PROCLEIX® HC+ Luminometer
Manufacturer
| MGM Instruments |
|---|---|
Dimensions (WxDxH)
| 18" x 29.5" x 7.5" (45.7 cm x 67.3 cm x 19.1 cm) |
Weight
| 50.7 lbs (23 kg) |
Workspace
| Room for PC and optional printer. Chamber lid opens at the top of the instrument. Power switch is recessed at the back. Fluids are refilled on the right side. Keypad is at the front. |
Voltage Requirements
| 100 - 140 VAC or 220 - 250 VAC 50 - 60Hz UPS: Must be capable of producing 800 VA for at least 7 min. |
Power
| 60 VA |
Operating Conditions
| Specifications below are specific to the instrument and are superseded by assay requirements. |
Temperature
| 15º to 30ºc |
Humidity
| 10% - 80% |
Altitude
| Up to 2000 m |
Consumables
| Reagents: Auto Detect I and II for assay processing, SysCheck for QC DI water is needed for bi-weekly warm water flush |
Purified Water
| System fluid is degassed, DI or distilled water (type III reagent grade [CAP/NCCLS]) |