TECAN® GENESIS® Robotic Sample Processor

System Capacity*
Continuous Onboard Capacity 1000 tubes
Assay Worklist 500 tubes
Sample Carousel 180 tubes in 9 sample racks
Hardware features
Tip configuration
4 or 8 tips; various combinations of different tip types possible
Tip variety
Standard (Teflon®-coated stainless steel), ceramic, short/long low- volume, disposable tips with or without filter (10 / 200 / 1000 µl)
Tip movement
Independent Z, equidistant Y (9 to 38 mm between tips), resolution of 0.1 mm in all three axes (X, Y, Z)
Syringe sizes
250 / 500 / 1000 / 2500 / 5000 µl
Liquid-handling features
Standard volume
3 - 5000 µl (standard tips), 3 - 1000 µl (disposable tips)
Low-volume option
0.5 - 250 µl (low-volume tips)
Password protection
Three password levels: user, integrator and administrator
Pipetting precision (CV)
Volume Standard tips Disposable tips
Standard volume*
200µl 1000µl
10µl < 1.5% < 1.5% ­
100µl < 0.4% < 0.4% < 0.4%
* typical results using pipetting water in single pipetting mode under optimized conditions (1000 µl syringe). [guaranteed for standard and disposable tips: 10 µl < 3 %, 100 µl < 0.5 %]
Pipetting precision (CV)
Volume Low-volume Standard tips Low-volume Disposable tips
Low volume*
1µl < 5 % < 7 %
10µl < 1.0% < 1.0% -
* typical results using pipetting water in single pipetting mode under optimized conditions (250 µl syringe). [guaranteed for standard and disposable tips: 1 µl < 10 %, 10 µl < 3 %]
Safety features
Liquid level detection
Down to 50 µl of conductive liquid can be detected in round-bottom microplates on standard carriers
Multiple liquid detection/
clot detection
Part of integrated liquid detection (ILID)
Password protection
Three password levels: user, integrator and administrator
Software features
Operating software
LOGIC and Gemini
Operating system
Windows™ NT or Windows™ 2000
Dimensions
RSP 100 RSP 150 RSP 200
height 830 mm / 32,7" 830 mm / 32,7" 830 mm / 32,7"
width 1051 mm / 41.4" 1426 mm / 56,1" 2026 mm / 79.8"
depth 780 mm / 30,7" 780 mm / 30,7" 780 mm / 30,7"
Power requirements
Power: 800 VA, voltage: 100-240 VAC, frequency: 50/60 Hz
Operating conditions
Temp.15-35º C/59-95F, rel.hum. 5-80% (30º C/86F or below)
TECAN and GENESIS are trademarks of TECAN Group AG
Teflon is a trademark of DuPont
Windows is a trademark of Microsoft Corp.