PROCLEIX® TIGRIS® System

System Capacity*
Continuous Onboard Capacity 1000 tubes
Assay Worklist 500 tubes
Sample Carousel 180 tubes in 9 sample racks
Footprint Dimensions (W x D)
PROCLEIX TIGRIS System 10.7 x 8 ft (3.25 x 2.44 m)**
Components Dimensions (W x D x H) Weight
PROCLEIX TIGRIS System 10.7 x 8 ft (3.25 x 2.44 m)** 1291 lbs (581 kg)
Analyzer 69 x 36 x 72 in (175.3 x 91.4 x 183 cm) 85 lbs (38 kg)
Computer Workstation 33 x 36 in (84 x 91.4 cm) 85 lbs (38 kg)
UPS 7 x 8.8 x 17.1 in (18 x 22 x 44 cm) 102 lbs (46 kg)
Vacuum Pump 12 x 16.5 x 14 in (31 x 42 x 36 cm) 35 lbs (16 kg)
Reagent Prep Incubator (RPI) 23 x 22.25 x 16.5 in (58 x 56 x 42 cm) 70 lbs (32 kg)
Environmental Requirements
Room Temperature 15º- 25ºC
Humidity 20% to 85%
Heat Dissipation 773W (2637 BTU/Hr) during steady operation
Electrical Requirements
Analyzer 200-240 VAC, 50-60 Hz, 30A (dedicated circuit required)
RPI US 120V AC, 60 Hz, 650 Watts
Sample Tube Bar Code Specifications
Code 39, Code 128 (ISBT 128), Codabar
Sample Size Tubes
12 x 75 mm, 13 x 75 mm, 13 x 100 mm, 16 x 75 mm, 16 x 100 mm
Operating System
Windows® XP
Type I Reagent-Grade Water
Special plumbing (e.g. direct water connection and/or floor drain) not required. Self-contained waste collection.
PROCLEIX is a trademark of Novartis Vaccines and Diagnostics, Inc.
TIGRIS is a trademark of Gen-Probe Inc.