PROCLEIX® TIGRIS® System
| System Capacity* | |
|---|---|
| Continuous Onboard Capacity | 1000 tubes |
| Assay Worklist | 500 tubes |
| Sample Carousel | 180 tubes in 9 sample racks |
| Footprint | Dimensions (W x D) |
|---|---|
| PROCLEIX TIGRIS System | 10.7 x 8 ft (3.25 x 2.44 m)** |
| Components Dimensions (W x D x H) | Weight | |
|---|---|---|
| PROCLEIX TIGRIS System | 10.7 x 8 ft (3.25 x 2.44 m)** | 1291 lbs (581 kg) |
| Analyzer | 69 x 36 x 72 in (175.3 x 91.4 x 183 cm) | 85 lbs (38 kg) |
| Computer Workstation | 33 x 36 in (84 x 91.4 cm) | 85 lbs (38 kg) |
| UPS | 7 x 8.8 x 17.1 in (18 x 22 x 44 cm) | 102 lbs (46 kg) |
| Vacuum Pump | 12 x 16.5 x 14 in (31 x 42 x 36 cm) | 35 lbs (16 kg) |
| Reagent Prep Incubator (RPI) | 23 x 22.25 x 16.5 in (58 x 56 x 42 cm) | 70 lbs (32 kg) |
| Environmental Requirements | |
|---|---|
| Room Temperature | 15º- 25ºC |
| Humidity | 20% to 85% |
| Heat Dissipation | 773W (2637 BTU/Hr) during steady operation |
| Electrical Requirements | |
|---|---|
| Analyzer | 200-240 VAC, 50-60 Hz, 30A (dedicated circuit required) |
| RPI | US 120V AC, 60 Hz, 650 Watts |
| Sample Tube Bar Code Specifications |
|---|
| Code 39, Code 128 (ISBT 128), Codabar |
| Sample Size Tubes |
|---|
| 12 x 75 mm, 13 x 75 mm, 13 x 100 mm, 16 x 75 mm, 16 x 100 mm |
| Operating System |
|---|
| Windows® XP |
| Type I Reagent-Grade Water |
|---|
| Special plumbing (e.g. direct water connection and/or floor drain) not required. Self-contained waste collection. |